Product Features | ||||
1 | Design of pure aluminum lamp body frame structure with ABS base, Aluminum Oxidation, ABS anti-UV coating. can prevent oxidation and discoloration under ultraviolet radiation. | |||
2 | Fold-able lamp bracket and 360-degree universal wheels make it convenient for storage and movement. | |||
3 | 0-180 degree adjustable lamp holder maximizes the disinfection area and sanitizing efficiency, nowhere to hide and run for the virus. | |||
4 | High purity quartz glass, strong UV penetration, 99.99% kill virus and bacteria, long lifetime. | |||
5 | Three-time settings, 15min, 30min, and 60min.Automatically shutdown after finish disinfection. | |||
6 | With Intelligent motion sensing function, it will automatically turn off when people entering the sensing area distance of 3-5 meters, sensitive sensing, no longer needto worry about children running around. | |||
7 | No radiation effect on people when the distance is out of 3-5 meters, just do not look directly at the light | |||
8 | Intelligent sensing technology, humanization design. 10 seconds delay starts to ensure the operator’s health. Motion sensing function will activate after turning on the lamp for 10 second | |||
9 | Optional accessory-protective guard in stainless steel | |||
10 | Optional accessory-replacement of tube for every wattage | |||
11 | Application Area: Offices, Kindergarten, school, factory, hospital, gym, hotel, restaurant, etc. |
Product specification for the Ultra Violet area disinfection system
Item Number | UVMS-UVC-80W | UVMS-UVC-120W | UVMS-UVC-150W | UVMS-UVC-200W |
Wattage | 80W | 120W | 150W | 200W |
PF | >0,5 | |||
Input Voltage | AC100V-120V or AC200V-240V | |||
Timing mode | Wireless remote control timing function | |||
Remote control mode | Wireless wall access remote control | |||
Delay range | 60 seconds delayed start | |||
Power cable length | 2 meter | |||
Lamp Current | 550 | 660 | 780 | 800 |
Base | 2G11 | 2G11 | 2G11 | 2G11 |
Base material | ABS | ABS | ABS | ABS |
UV Tube size | 411 | 411 | 533 | 533 |
Tube diameter | Ø 17mm | |||
UV intensity@1 meter | 260uw/cm² | 320uw/cm² | 360uw/cm² | 410uw/cm² |
Product size | 250*250*690mm | 250*250*690mm | 250*250*810mm | 250*250*810mm |
Weight (g) | 3250 | 3280 | 3700 | 3700 |
Lifetime Tube | 8000 hrs | 8000 hrs | 8000 hrs | 8000 hrs |
2000hrs UVC output maintenance | ≥80% | |||
Dielectric strength (V) | AC1500V | |||
Anti Electric Shock | CLASS 1 | |||
Implement standard | EN 61199 - EN 60901 | |||
Working condition | -10℃ ~ +45℃、10% ~ 90% R.H. | |||
Storage condition | -10℃ ~ +45℃、0% ~ 80% R.H. | |||
IP | IP20 Indoor | |||
Package size | 300*300*710mm | 300*300*710mm | 300*300*830mm | 300*300*830mm |
Netto weight | 3,25KG | 3,28KG | 3,70KG | 3,70KG |
Gross weight | 4,30KG | 4,30KG | 4,80KG | 4,80KG |
Outer carton size | 310*310*720mm (1pcs) | 310*310*720mm (1pcs) | 310*310*840mm (1pcs) | 310*310*840mm (1pcs) |
Recommended sterilization time | |||||
Area | Wattage | Minutes | Area | Wattage | Minutes |
30-50 square meters | 80W | 30 | 50-70 square meters | 80W | 60 |
50-70 square meters | 120W | 30 | 50-70 square meters | 200W | 15 |
70-100 square meters | 120W | 60 | 70-100 square meters | 200W | 30 |
100-120 square meters | 200W | 60 | 120-150 square meters | 200W | 90 |
Test Result(s) SGS: Validation of disinfection efficacy of sterilization and disinfection instruments* Test Method: Technical Standard for Disinfection - 2.1.5.4 | |||||
Test organism | Test groups | Average number of positive controls(cfu/pcs) | Average number of testing groups (cfu/pcs) | Killing rate % | Sterilization logarithm |
Escherichia coli | 1 | 1.6×10^6 | 1.4×10^4 | 99.12 | 2.05 |
2 | 1.5×10^6 | 1.3×10^4 | 99.13 | 2.06 | |
3 | 1.7×10^6 | 1.4×10^4 | 99.18 | 2.08 | |
Staphylococcus aureus ATCC 6538 | 1 | 4.2×10^6 | 2.4×10^4 | 99.43 | 2.24 |
2 | 4.0×10^6 | 2.3×10^4 | 99.42 | 2.24 | |
3 | 3.4×10^6 | 2.2×10^4 | 99.35 | 2.19 | |
Candida albicans ATCC 10231 | 1 | 3.5×10^6 | 1.9×10^4 | 99.46 | 2.26 |
2 | 4.0×10^6 | 2.0×10^4 | 99.50 | 2.30 | |
3 | 3.7×10^6 | 1.8×10^4 | 99.51 | 2.31 | |
Remark | |||||
1. Sample pretreatment: The glass was used as the carrier of the microbial sheet. During the test, the microbial sheet was placed 2m away from the UV lamp. Treatment time was 1h. 2. * The test was carried out by external laboratory assessed as competent. | |||||